Skip to content
View Categories

Serum Albumin Test

2 min read

INTRODUCTION

Albumin is the most abundant protein in blood plasma. Measurement of serum albumin provides important information about liver and kidney function. This document provides the procedure for colorimetric quantification of albumin using bromcresol green dye. Standardization of this technique aims to ensure accurate, reliable test results for optimal clinical interpretation.

Test Catageory
فئة التحليل

Clinical Chemistry

Refrence Books
كتب مراجع

1. Bishop’s Clinical chemistry

Skills Required
المهارات المطلوبة

1. Pipetting
2. Colorimetry

Video Resources
مصادر فيديو

عربي
English

Purpose of the Test

To determine the level of albumin in serum samples.

Principle

Albumin in the sample binds with the dye bromcresol green (BCG) to form a colored complex. The intensity of the color formed is directly proportional to the albumin concentration.

Equipments

1. Semi-autoanalyzer
2. Cuvettes
3. Micropipettes
4. Timer
5. Spectrophotometer
6. Albumin Reagent (Contains succinate buffer and bromcresol green dye)

Sample Type

Serum or plasma is acceptable for the Test

Procedure

1. in 3 clean dry tubes add reagents and sample as mentioned in the Table below

BlankSTDTEST
Albumin Reagent1000 μl1000 μl1000 μl
STD5 μl
Sample5 μl

Procedure

2. Mix Well and incubate for 5 mins at room Tempreature or 5 min @ 37c’
3. using the colorimeter or spectrophotometer read the absorbance against blank at filter wavelenght 620 nm
4. then do the calculations based on the following formula

Albumin = Absorbance of test / Absorbance of standard x Conc of standard
The STD Conc is 5 g/dl or may vary depending on the reagent brand.
Reference range: 3.5 – 5.5 g/dL
Normal Range

Quality Control:

1. Run two levels of control with each batch of samples.
2. Results should fall within established confidence limits.
3. Review QC results and take corrective action if needed.

Warning :

1. Contamination can cause erroneous results.

Quick Notice

M. Sulieman

mohammad@mlsgaate.com

Check protein reagent expiration and storage conditions. Inspect instrument calibration. Follow SOP for total protein testing.
تحقق من تاريخ انتهاء صلاحية معاير الألبيومين . فحص الجهاز للتأكد من عدم وجود أخطاء. شغّل عينات التحكم قبل فحص العينات.