Direct Bilirubin Test

1 min read

INTRODUCTION

Measurement of direct or conjugated bilirubin is important for the diagnosis of hepatobiliary disease and liver function assessment. This Procedure provides the procedure for photometric determination of direct bilirubin using diazotized sulfanilic acid. Standardization of this analytical technique aims to ensure accurate, reliable test results for optimal clinical interpretation and medical decision making.

Test Catageory
فئة التحليل

Clinical Chemistry

Refrence Books
كتب مراجع

1. Bishop’s Clinical chemistry

Skills Required
المهارات المطلوبة

1. Pipetting
2. Colorimetry

Video Resources
مصادر فيديو

عربي
English

Purpose of the Test

To determine the level of direct (conjugated) bilirubin in serum samples.

Principle

Direct bilirubin reacts with diazotized sulfanilic acid to form a blue colored azobilirubin complex. The intensity of the color formed is directly proportional to the concentration of direct bilirubin in the sample.

Equipments

1. Semi-autoanalyzer
2. Cuvettes
3. Micropipettes
4. Timer
5. Spectrophotometer
6. Total Bilirubin Reagent (Contains sulfanilic acid and hydrochloric acid)

Sample Type

Serum preferred for the Test

Procedure

1. in 3 clean dry tubes add reagents and sample as mentioned in the Table below

Sample BlankSample
Direct Bilirubin RGt1000 μl1000 μl
Activator / Substrate50 μl / 1 drop
Sample100 μl100 μl

Procedure

2. Mix Well and incubate for 10 mins at room Tempreature or 5 min @ 37c’ (protected from light.)
3. using the colorimeter or spectrophotometer read the absorbance of Sample against Sample blank at filter wavelenght 540 nm
4. then do the calculations based on the following formula

Total Bilirubin = (Abs.Test – Abs. Sample Blank) X (Factor)
Reference range: Adult: 0.0 – 0.3 mg/dL Newborn: 1 – 2 mg/dL
Normal Range

Quality Control:

1. Run two levels of control each day of testing.
2. Control results should fall within established range.

Warning :

1. Hemolysis causes falsely Increased results.
2. Photodegradation can falsely lower results.

Quick Notice

M. Sulieman

mohammad@mlsgaate.com

Check reagent expiration. Inspect instrument for errors. Run controls before testing samples.